Pressure 7.50, Elapsed – Fluke Biomedical IDA-5 User Manual
Page 29

Infusion Device Analyzer
Occlusion Tests
19
When you select Start, the Channel Occlusion screen in Table 13 shows
measurement data as the test continues. Select Graph to show the test data as
a graph.
Table 13. Occlusion Screen in Active Mode
C h a n n e l 1 O c c l u s i o n
Pressure 7.50
psi
Elapsed
Time 00:09:10
Peak Pressure
8.04
psi
Time of Peak
00:08:23
Graph
End
Press ESC for Status Screen
gir14.eps
Measurement Description
Pressure
Shows the amount of pressure and is set by the Operator or the User Defaults.
Elapsed Time
Time since the test was started.
Peak Pressure
Highest pressure sensed since the test started.
Time of Peak
The time that highest pressure was sensed.
The Occlusion Graph screen for the same test is shown in Figure 7. The same
numerical data is shown.
C h a n n e l 1 Oc c l u si o n G r a p h
Inst psi
7.50
Peak psi
8.04
Elapsed Time
00:09:10
Peak
00:07:23
10
psi
0
0 Min 30
View Detail
Press ESC for Status Screen
gir15.eps
Figure 7. Occlusion Graph Screen
Select View Detail to see the Occlusion Detail screen shown in Table 14.