Measuring Flow and Volume
Measuring Volume
6
6-5
High Flow Sensor
To Ventilator
edg24f.eps
Figure 6-3. Measuring High Flow
OTHER
MEASUREMENTS
WAVEFORM
SCREENS
edg25f.eps
Figure 6-4. Accessing High-Flow Measurements
Measuring Volume
The Analyzer integrates (sums) the high- or low-flow signal over a period of time to
obtain volume. The Analyzer measures volume in units of Liters, CF (cubic feet), or mL
(milliliters). Maximum low-flow volume is ± 60 l, and maximum high-flow volume is >
± 60 l. All correction modes and other measurement accuracy considerations applicable
to flow measurement apply for volume measurements. To make a volume measurement,
follow these steps:
1. Connect the Analyzer for low-flow or high-flow measurement as shown in Figure 6-1
or 6-3, respectively.
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)