High-flow port, Low-flow port – Fluke Biomedical VT Mobile User Manual
Page 101
Appendices
Specifications
B
B-3
High-Flow Port
Maximum Flow Rate (absolute value):
200 lpm
Operating Flow Range:
± 150 lpm
Accuracy:
± 3 % of reading or ± 2 % of range, whichever is
greater
Floor for Absolute Accuracy:
25 lpm
Resolution:
0.01 lpm
Frequency Response:
> 25 Hz or t
10-90
< 40 ms
Sample Rate:
100 Hz
Dynamic Resistance:
< 2 cmH
2
O @ 60 lpm
Low-Flow Dropout:
2.5 lpm
Breath-Detect Threshold:
4 lpm (user selectable)
Volume Range:
> ± 60 l
Tidal Volume Accuracy:
± 3 % or reading or ± 20 ml, whichever is greater
Low-Flow Port
Maximum Flow Rate (absolute value):
35 lpm
Operating Flow Range:
± 25 lpm
Accuracy:
± 3 % of reading ± 1% of range, whichever is greater
Floor for Absolute Accuracy:
3 lpm
Resolution:
0.01 lpm flow > 1 lpm
Freq. Response:
> 25 Hz or t
10-90
< 40 ms
Sample Rate:
100 Hz
Dynamic Resistance:
< 2.5 cmH
2
O @ 5 lpm
Low-Flow Dropout:
0.24 lpm
Breath-Detect Threshold:
1 lpm (user selectable)
Volume Range:
± 60 l
Tidal Volume Accuracy:
± 3 % of reading or ± 5 ml