Airway pressure screen, Volume screen – Fluke Biomedical VT Mobile User Manual
Page 32

VT MOBILE
Operators Manual
3-8
Press G to change the units of measure for the pressure screen. Possible units of
measure for this screen include:
•
mbar:
millibar = .001 Bar
•
bar:
14.7
PSIG
(Barometric)
•
mmHg:
millimeters of mercury
•
inHg:
inches of mercury
•
cmH
2
O:
centimeters of water
•
inH
2
O:
inches of water
•
PSIG:
pounds per square inch
•
atm: atmospheres
•
kPa: kilopascals
Press F (Rescale) to change the scale on the plot of data to a more appropriate
window of values.
Note
Rescale uses the current Min/Max values to determine plot limits.
Airway Pressure Screen
Press L as required to access the Airway Pressure Screen (Paw), which displays the
reading and waveform measured in the high- or low-flow port with the Analyzer
connected between a ventilator and a test lung. The following is a sample Airway
Pressure Screen:
ede35s.bmp
Volume Screen
Press O to access the Volume Screen. The volume screen displays the reading and
waveform for the volume signal. The volume signal is the integral of the flow signal. A
sample volume screen appears below:
ede36s.bmp
Press G (Units) to change the units of measure for the volume screen. Possible units
of measure for this screen include:
•
Liters Liters
•
mL
Milliliters
•
CF
Cubic Feet