Screens, Ventilator parameters screens – Fluke Biomedical VT Mobile User Manual
Page 28
VT MOBILE
Operators Manual
3-4
Screens
Ventilator Parameters Screens
The Analyzer calculates 16 parameters for every breath. You can display all parameters
individually or most parameters in groups of four. Parameters update following every
breath. Units of measure for breath parameters are fixed. Table 3-2 shows parameters and
groupings, along with abbreviations and units of measurement.
Table 3-2. Ventilator Parameters
Parameter Units
Inspiratory and Expiratory Tidal Volume (VT)
ml
Expiratory Minute Volume (MV)
lpm
Peak Inspiratory Flow (PIF)
lpm
Peak Expiratory Flow (PEF)
lpm
Peak Inspiratory Pressure (PIP)
cmH
2
O
Positive-End Expiratory Pressure (PEEP)
cmH
2
O
Mean Airway Pressure (MAP)
cmH
2
O
Inspiratory Pause Pressure (IPP)
cmH
2
O
Breath Rate (Rate)
bpm
Inspiratory Time (Ti)
s
Expiratory Time (Te)
s
Inspiratory, Expiratory Time (I:E) Ratio
s:s
Oxygen Concentration
% O
2
Temperature (optional)
°
C (selectable)
Relative Humidity (optional)
% RH
Barometric Pressure
mmHg (selectable)