Preparing for Operation
Setting Up
2
2-7
The Analyzer Zero Warn is set to ON at the factory to remind you to zero out any offset
in the Flow measurements. The Zero Warn screen reminds you at power on, after the first
five minutes, and then after each 30-minute period. After the power-on zero warning, you
can choose to Zero or Cancel when the reminder appears; either action returns the
Analyzer to the same function.
W
Caution
Setting Zero Warn to OFF may result in introduction of offset
into the Flow and Volume measurements and cause errors in
the assessment of those measurements against the medical
device under test specification. Fluke Biomedical strongly
recommends that you leave Zero Warn ON.
Turn this warning on or off using the following procedure:
1. Press
D
Æ E Æ C (to ‘Zero Warn’).
2. Press E Æ C (to select ON or OFF).
3. Press E to activate your selection.
Setting the Time
You can set hours and minutes. Enter the time in the HHMM or HMM format, where HH
or H is the hours in 12-hour format and MM is the minutes (00 through 59).
To set the time, follow these steps:
1. Press
D
Æ C Æ E Æ C Æ E to access the time setting screen.
2. Enter the hour (HH or H) with the text entry keys.
3. Press
G
.
3. Enter the minutes (MM) with the text entry keys.
4. Press F to select P (PM) or A (AM).
5. Press E.
Setting the Date
To set the date, follow these steps:
1. Press
D
Æ CÆEÆ E to access the date setting screen.
2. Enter the month (MM) or day (DD), depending on the date format, with the text entry
keys.
3. Press G.
4. Enter the day (DD) or month (MM) with the text entry keys.
5. Press G.
6. Enter the year (YYYY) with the text entry keys.
7. Press E.
The Analyzer displays the date with the following information:
•
MM is the number between 1 and 12 corresponding to the month.
•
DD is the number between 1 and 31 corresponding to the day of the month.
•
YYYY is the year.