Oxygen measurement, Barometric pressure measurement, Temperature and relative humidity – Fluke Biomedical VT Mobile User Manual
Page 102
![background image](https://www.manualsdir.com/files/785355/content/doc102.png)
VT MOBILE
Operators Manual
B-4
Oxygen Measurement
Range:
0 to 100 %
Accuracy:
± 2 % full-scale output
Resolution:
0.1 % O
2
Frequency Response:
> 15 s (t
10-90
)
Sample Rate:
100 Hz
Sensor Technology:
Galvanic Fuel Cell
Calibration:
Allows user calibration using air and 100 % O
2
Notes:
• Automatic partial pressure compensation for barometric and airway pressure changes
• Recommended interval for changing the Oxygen Sensor is one year. However, sensor may last
longer. During user calibration of the sensor, the Analyzer can detect if the sensor needs
replacement.
• Maxtec Oxygen Sensor distributed by Fluke Biomedical
Barometric Pressure Measurement
Range:
8 to 18 psia (400 to 900 mmHg)
Accuracy:
± 2 % of reading
Resolution:
0.1 mmHg
Frequency Response:
< 5 s (t
10-90
)
Sample Rate:
N/A
Calibration:
Not required; however, device allows user calibration of offset.
Temperature and Relative Humidity
Temperature Relative
Humidity
Resolution:
0.1
°
0.1 %
Range:
0 – 50
°
C
10 to 95 %
Accuracy:
± 1
°
C
± 10 % RH
Units:
°
C,
°
F,
°
K
%