Measurement parameter specifications, Display screen, Operational modes – Fluke Biomedical VT Mobile User Manual
Page 103: Gas types, Serial communications specifications
Appendices
Specifications
B
B-5
Peak Inspiratory Pressure (PIP)
0.1 cmH
2
O
± 120 cmH
2
O
± 3 % or 2 cmH
2
O
Inspiratory Pause Pressure (IPP)
0.1 cmH
2
O
± 120 cmH
2
O
± 3 % or 2 cmH
2
O
Mean Airway Pressure (MAP)
0.1 cmH
2
O
± 80 cmH
2
O
± 3 % or 1 cmH
2
O
Positive-End Expiratory Pressure (PEEP)
0.1 cmH
2
O
- 5 to 40 cmH
2
O ± 3 % or 1 cmH
2
O
Peak Expiratory Flow (PEF)
0.01 lpm
0 to 150 lpm
± 3 % or 1 lpm
Peak Inspiratory Flow (PIF)
0.01 lpm
0 to 150 lpm
± 3 % or 2 lpm
Inspiratory Time (Ti)
0.01 s
0 to 60 s
0.5 % or 0.02 s
Expiratory Time (Te)
0.01 s
0 to 90 s
0.5 % or 0.01 s
Display Screen
64 x 128 pixels, reflective LCD, blue on yellow
Operational Modes
Standalone
Controlled by VT for Windows PC Software (version 2.01.00 or higher)
Gas Types
Air, O
2
, Heliox, CO
2,
N
2
, N
2
O, N
2
O bal O
2
, He bal O
2
, N
2
bal O
2
User
Serial Communications Specifications
4-pin miniature RS-232 serial port located on upper-left side of the Analyzer
RS232 compatible with the VT for Windows software application (version 2.01.00 or
higher.)
Measurement Parameter Specifications
Parameter Resolution
Range
Accuracy
Inspiratory and Expiratory Tidal Volume (VT)
0.1 ml
> 10 l
± 3 %
Expiratory Minute Volume (MV)
0.001 lpm
0 to 60 l
± 3 %
Breath Rate (BPM)
0.1 bpm
2 to 150 bpm
± 1 %
Inspiratory, Expiratory Time (I:E)
0.01 s
0.25 to 9.99 s
± 2 % or 0.1 s