Measuring Low Pressure
Measure low pressure in the range: -20 to 120 cmH
2
O. Never exceed an absolute
maximum of 5 psi (0.345 bar).
Refer to Figure 5-1 while following these steps:
1. Press R and follow on-screen instructions.
2. At the Analyzer, connect the Low-Pressure Adapter to the Flow/Low-Pressure Port.
3. Press L in the OTHER MEASUREMENTS or WAVEFORM SCREENS section.
4. Verify that the screen displays Low Pressure (Plo) or LPress. If the High-Pressure
Adapter is also connected, you may need to press L a second time.
5. Connect the Low-Pressure Adapter to the source.
6. Apply pressure.
7. Successively press G (Units) to select a pressure unit in the sequence: PSI Æ kPa
Æ Bar Æ mBar Æ ATM Æ inH2O Æ InHg Æ cmH2O Æ mmHg.
If the waveform is too small or too large, press F (Rescale) to view the waveform
within optimized limits.
F1
F2
Back
ENTER
Setup
Memor
y
Memor
y
VENTILA
TOR P
ARAM
ETERS
FLO
W/
VO
L
FLO
W/
VO
L
TREND
TEST
ZER
O
SENSORS
MORE
PRESSURE
PRESSURE
PRESSURE
TIM
ING
VOLUM
E
FLO
W
W
AVEFORM
SCREENS
OTHER MEASUREMENTS
SPECIAL FUNCTIONS
%RH
O
2
_ , @
_ , @
11
22
AB
C
AB
C
55
JK
L
JK
L
44
GH
I
GH
I
66
MN
O
MN
O
88
00
TU
V
TU
V
77
PQ
RS
PQ
RS
99
WX
YZ
WX
YZ
33
DE
F
DE
F
GAS FLOW ANALYZER
GAS FLOW ANALYZER
VT
M
OB
ILE
VT
M
OB
ILE
Low
Pressure
Adapter
edg08f.eps
Figure 5-1. Measuring Low Pressure