Page 29: Oxygen screen

Using the VT MOBILE
Screens
3
3-5
Viewing a Single Parameter
To view single parameters, press K, L, or M repeatedly. If you are already
viewing a multi-parameter screen, press G (Single). The following is an example of
the PIP single parameter screen:
ede26s.bmp
Viewing a Group of Parameters
From an individual parameter screen, press G (All) to display the related screen of
four parameters. The groups are Pressures, Vol/Flow, and Timing. The following is an
example of the Pressures multi-parameter screen:
ede27s.bmp
Viewing All Parameters
You can view all 16 parameters on one screen when using the Analyzer connected to a
PC running the VT for Windows software.
Oxygen Screen
Press P to access the Oxygen Screen. The sample oxygen screen shown below
displays the signal and statistics for the oxygen sensor.
ede28s.bmp
Press F (Clear) to reset the statistics and start a new reading.
Press G (Calibrate) to begin the calibration routine for the oxygen sensor; you will
find complete instructions later in this chapter.