Fluke Biomedical VT Mobile User Manual
Page 37
Using the VT MOBILE
Using Trend Test
3
3-13
8. When both Low and High limit values are correct, press E again to advance to
the Start Trend Test screen.
ede46s.bmp
9. Before starting the test, ensure that the ventilator is on, has reached steady-state
conditions, and that the displayed breath parameter is as expected. Press G (Start)
to begin the test.
10. The Analyzer displays the Trend Test screen while the trend test is in progress. The
trend test screen shows the parameter being tested, the current value, the upper and
lower limits, and the number of events (readings falling outside the limits.)
ede47s.bmp
11. To terminate the test, press G (Stop). The Trend Summary screen appears. If there
are no events, the Trend Summary screen is not available.
ede48s.bmp
ede49s.bmp
12. Alternatively, press G (More) to display related parameters.
ede50s.bmp
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)