beautypg.com

1 general product information – Bronkhorst DeviceNet slave interface User Manual

Page 5

background image

BRONKHORST

®

 

Page 5   

 

 

 

 

DeviceNet interface 

 

 

 

9.17.026 

1 GENERAL PRODUCT INFORMATION 

1.1 I

NTRODUCTION

 

 
The  DeviceNet  interface  offers  a  direct  connection  to 
DeviceNet  Networks  for  Bronkhorst

1)

  digital  mass‐

flow/pressure  meters/controllers  according  to  the  Mass 
Flow  Controller  Profile  specified  by  the  ODVA.  This 
manual  is  limited  to  the  description  of  the  interface 
between  the  DeviceNet  Mass  Flow  Controller  with  a 
master device.  
 
This  manual  will  explain  how  to  install  a  Bronkhorst  instrument  to  your  DeviceNet  system.  It  only  consists  of  that 
information which is needed most.  
 
There is no mutual communication between DeviceNet slaves, only between master and slave. Each slave should have 
its  own  unique  MAC  ID  on  the  network,  otherwise  there  is  no  communication  possible.  Setting  MAC  ID  can  be 
performed by either: 

 Master configuration software 
 Bronkhorst tooling software: FlowFix (on Multibus documentation/software tool CD) 

This programme is able to communicate with the instrument via RS232 using a special 
cable. If you don’t have such a cable, ask your local sales representative. 

 Button (+ LED's) on top of the instrument 
 Rotary switches on the side of the instrument (if present). 

 
 

 

1)

 Bronkhorst: 

This includes Bronkhorst High‐Tech B.V. , Bronkhorst Cori‐Tech B.V. and 

 

 

 M+W Instruments GmbH. 

 

1.2 M

ULTIBUS TYPES

 

 
In 2000 Bronkhorst developed their first digital instruments according to the “multibus” principle. The basic pc‐board 
on  the  instrument  contained  all  of  the  general  functions  needed  for  measurement  and  control,  including  alarm, 
totalizing and diagnostic functions. It had analog I/O‐signals and also an RS232 connection as a standard feature. In 
addition  to  this  there  is  the  possibility  of  integrating  an  interface  board  with  DeviceNet™,  Profibus‐DP

®

Modbus  , 

FLOW‐BUS  or  EtherCAT  protocol.  The  first  generation  (MBC‐I)  was  based  on  a  16  bit  Fujitsu  controller.  It  was 
superseded in 2003 by the Multibus type 2 (MBC‐II). This version was also based on 
the  16  bit  Fujitsu  controller  but  it  had  several  improvements  to  the  MBC‐I.  One  of 
them is the  current steering  of the valve. It reduced  heat production and improved 
control  characteristics.  The  latest  version  Multibus  controller  type  3  (MBC3)  is 
introduced in 2011. It is build around a 72MHz 32 bit NXP ARM controller. It has AD 
and  DA  controllers  on  board  which  makes  it  possible  to  measure  noise  free  and 
control valves without delays. The internal control loop runs 6 times faster compared 
to the MBC‐II therefore control stability has improved significantly. It also has several 
improved  functions  like  reverse  voltage  protection,  inrush  current  limitation  and 
overvoltage protection.  
 
 
MBC3
 instruments can be recognised by the “MBC3” placed on lower left side  
of the instrument label (see example).  
 

 

Information about DeviceNet can be found at the website of de ODVA 
organisation. 

www.odva.org