12 TROUBLESHOOTING
12.1 LED
INDICATIONS
LED indications can be very useful in case of problems with the instrument.
The green LED is normally used for instrument status indication, like normal operation or special function mode. For
DeviceNet it is also possible to show that the instrument is in abort state and idle state. The red LED is normally used
to for error/warning indication (how longer the flash, how greater the trouble).
12.1.1 LED indications mode (MBC‐II and MBC3)
Led
Time
Indication
Green
off
Continuous Power‐off or program not running
on
Continuous Normal running/operation mode
Short
flash
0.1 sec on
2.0 sec off
Initialization mode Abort state
Secured params can be changed
normal
flash
0.2 sec on
0.2 sec off
Special function mode
Instrument is busy performing any special function.
E.g. auto‐zero or self‐test
long flash 2.0 sec on
0.1 sec off
idle state
Red
off
Continuous No error
Short
flash
0.1 sec on
2.0 sec off
Minor communication error
normal
flash
0.2 sec on
0.2 sec off
No bus power
long flash 2.0 sec on
0.1 sec off
Serious communication error
manual intervention needed
long flash 1.0 sec on
0.1 sec off
For special service purpose only
on
Continuous Critical error message.
A serious error occurred in the instrument.
Instrument needs service before further using.
Wink Mode
Green
Red
Green
Red
turn by turn
slow
wink
0.2 sec on
0.2 sec off
Wink mode
By a command send via FLOW‐BUS the instrument can “wink” with Led’s to indicate its
position in a (large) system
normal
wink
1.0 sec on
1.0 sec off
Alarm indication: minimum alarm, limit/maximum alarm; power‐up alarm or limit exceeded or
batch reached.
fast
wink
0.1 sec on
0.1 sec off
Switch‐released, selected action started