DeviceNet interface
9.17.026
9.4 V
IA
RS232:
OTHER PROGRAMS
It is also possible to read and or change station address or baud rate by means of any program via RS232 using the
COM‐port of your PC on 38.4 KBaud. This can be achieved using the FLOW‐BUS protocol.
The following table gives the parameters in proc 125 which may be used:
Parameter
Type
R/W
Init mode
Description
9
LONG
R/W
Soft init
Baud rate for field bus interface
10
CHR
R/W
Soft init
Field bus station address/MAC ID
More detailed information about the RS232 protocol (document 9.17.027) can be found at:
http://www.bronkhorst.com/en/downloads/instruction_manuals/
9.5 V
IA MICRO
‐
SWITCH AND
LED
S ON TOP OF INSTRUMENT
With the micro‐switch on top of the instrument it is possible to change and readout the settings for MAC‐ID and baud
rate. The LED’s will indicate the tens of the address with green flashes and the units with red flashes. For baud rate
indication both LED’s will flash.
See document 9.17.023 for a detailed description
This document can be found at:
http://www.bronkhorst.com/en/downloads/instruction_manuals/