9

2.8

Analog / digital operation

Analog operation

For analog operation refer to the “Hook-up diagram LIQUI-FLOW

TM

mini” or use

an RJ45 loose-end cable (7.03.419) to connect the required signals.

Digital RS232 operation

Digital operation over RS232 can be established

when using the following setup. Using a RS232

cable or a USB-RS232 converter with a PC will

allow you to use (free) Bronkhorst® software for

Windows, such as FlowDDE and FlowPlot. See

also section 3.4.

Digital RS485 operation

With digital operation over RS485 a bus-system with multiple instruments can be set up. See section 3.5 for possible

systems.

2.9

Multi-functional switch operation

Micro-switch operation (single channel versions only)

Using the micro-switch on the instruments, several actions can be monitored and

started. The green LED is used for status indication. The red LED is used for

errors/warnings/messages. The micro-switch can be used to start several

actions, such as auto-zero, restore factory settings and bus-initialisation actions,

if applicable. See specific zero-procedure below and section 3.3 for more details.

2.10 Purging

When applying pressure to the system, avoid pressure shocks in the system and increase pressure gradually. Also

decrease pressure gradually when required.

In systems for use with toxic or other dangerous fluids, purging for at least 30 minutes with an inert fluid is

absolutely necessary before use. After use with toxic or other dangerous fluids, complete purging is also required

before exposing the system to air.

Waste fluid

Make sure not to spill fluid on the instrument while purging.

Always keep the sensor exterior dry.

Warm-up time

Let the LIQUI-FLOW

TM

mini warm-up for at least 30 minutes for best accuracy. During warm-up period, fluid pressure