3.5.1.2 Modbus setup
3.5.1.3 RS232 stand alone setup
3.5.2
Software
When using a pc to communicate with LIQUI-FLOW
TM
mini instruments only the FLOW-BUS protocol is supported by Bronkhorst
software. When using Modbus operation, software from third parties, such as LabView, ModScan or a Modbus PLC must be used
to serve as Modbus master.
Note: a LIQUI-FLOW
TM
mini instrument set for RS485 FLOW-BUS or Modbus communication will not respond when
connecting to an RS232 configuration. When required press the micro-switch according to the procedure in section
3.3.1 at start-up to activate the “Configuration Mode”. The baud rate and bus type are set to 38K4 and RS232
FLOW-BUS (Propar).
3.5.3
Slave address, baud rate and parity setup
The LIQUI-FLOW
TM
mini instrument is configured as specified on order. If there is a need of changing any of the specified settings,
see the tables below for the supported configurations.
Mode:
Analog
Digital
Interface/medium:
-
RS232
RS485
Bus protocol:
-
Propar
FLOW-BUS
Modbus RTU
Modbus ASCII
Baud rate:
-
9600
19200
38400
57600
115200
187500
400000
9600
19200
38400
56000
57600
115200
9600
19200
38400
56000
57600
115200
Node address:
-
3…125
3…125
1…247
1…247
Parity:
-
None*
None*
None; Even; Odd
None; Even; Odd