beautypg.com

11 zeroing, 12 calibration, 13 supply pressure – Bronkhorst LIQUI-FLOW mini (from 01-07-2013) User Manual

Page 10: 14 measuring principle, Zeroing, Calibration, Supply pressure, Measuring principle

background image

LIQUI-FLOW™ mini

9.17.065

Bronkhorst High-Tech BV

10

may either be on or off.

2.11 Zeroing

The zero point of each instrument is factory adjusted. If required the zero point can be re-adjusted over RS232 or by

means of using the micro-switch. Procedure for zeroing by-micro switch:

Warm-up, pressure up the system and fill the instrument according to the process conditions.

Make sure no flow is going through the instrument by closing valves near the instrument.

Press micro-switch and hold it. After a short time the red LED will go ON and OFF, then the green LED will go ON.

At that moment release the micro-switch.

The zeroing procedure will start at that moment and the green LED will blink fast. The zeroing procedure waits for

a stable signal and saves the zero. If the signal is not stable, zeroing will take long and the nearest point to zero is

accepted. The procedure will take approximately 10 seconds.

When the indication is showing 0% signal and the green indication LED is burning continuously again, the zeroing

action was successful.

2.12 Calibration

Each LIQUI-FLOW

TM

mini instrument is factory calibrated. Bronkhorst High-Tech B.V. certifies that all instruments

meet the rated accuracy. Calibration is performed using measurement standards traceable to the standards of the

Dutch Metrology Institute (VSL). Calibration certificates are included in the shipment. When operated properly

(clean fluid, no pressure shocks, no vibrations, no thermal shocks, etc.), regular maintenance is not required.

However, periodical inspection, recalibration or verification of the accuracy may be subject to individual

requirements of the end-user.

2.13 Supply pressure

It is recommended to turn on power before applying pressure on the instrument and to switch off power after

removing pressure.

Turn on fluid supply gently. Avoid pressure shocks and bring the instrument gradually up to the level of the actual

operating conditions. Also switch off fluid supply gently.

2.14 Measuring principle

The liquid flow sensor operates on a principle of heat transfer by sensing the

temperature difference along a heated section of a capillary tube. The

temperature difference sensed by the upstream and downstream

temperature sensors on the capillary depends on the amount of heat

absorbed by the fluid flow. The temperature sensors are part of a bridge

circuit, as seen in the figure. The measured imbalance of the bridge, caused by

the flow in the capillary is linearised to the mass flow rate and amplified to the

desired signal level.

Thermal sensor in a bridge configuration

The transfer function between mass flow and sensor signal can be described by the equation:

V

signal

= K . c

p

. Ö

m

with:

V

signal

= sensor signal

K

= constant factor

c

p

= specific heat of the fluid

Ö

m

= mass flow

From this formula it can be concluded that for each different fluid (with heat capacity c

p

) a certain sensor signal represents a

different mass flow. Using accurate fluid data and a conversion factor it is possible to predict the mass flows for liquids other than

water.