2 fluid information, Fluid information – Bronkhorst LIQUI-FLOW mini (from 01-07-2013) User Manual

Page 21

background image

LIQUI-FLOW™ mini

© 2013 Bronkhorst High-Tech BV

Bronkhorst High-Tech BV

21

Value

Mode

Instrument action

Setpoint

source

Master

source

Slave

factor

slave factor /

100%

analog

input

18

RS232

Controlling

RS232

20

Valve steering

Setpoint is redirected to Valve Out with

controller idle

21

Analog valve

steering

Analog input is redirected to Valve Out with the

controller idle

22

Valve safe state

After power-up the control mode will be set to 'Analog input' or 'BUS/RS232', depending on the customer's default setting for

analog or digital operation. Except when the actual control mode setting is other than 0, 1, 9 or 18 the actual control mode setting

is maintained. For more information see parameter 'IOStatus', section 4.2.2.

i

www

For dual interface operation or slave factors, see document nr. 9.17.023: Operational instructions for digital

instruments. http://www.bronkhorst.com/files/downloads/

manuals_english/917023_operation_instructions_digital_instruments.pdf

4.1.2

Fluid information

The following parameters give information about the selected fluid range.

Fluid Number

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char

RW

0…7

24

1/16

0x0030/49

The ‘Fluid number’ is a pointer to the set of calbration parameters. Each selectable fluid has its own set of calibration parameter

values. Parameter value 0 = fluid 1 and parameter value 7 = fluid 8. Up to eight fluids can be stored in an instrument. Default value

= 0 (fluid 1).

Fluid Name

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char[10]

RW

-

25

1/17

0x8188…0x818C/33161…33165

This parameter consists of the name of the selected fluid number, e.g. ‘

Air

’.

Fluid Unit

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char[7]

RW

-

129

1/31

0x81F8…0x81FB/33273…33276

The ‘Fluid Unit’ can be read by parameter ‘Capacity Unit’. This parameter contains the unit in maximal 7 characters.

Fluid Capacity (@100%)

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

RW

±1E-10…

±1E+10

21

1/13

0x8168…0x8169/33129…33130

Capacity is the maximum value at 100% for direct reading in sensor readout units.

Fluid Capacity (@0%)

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

RW

±1E-10…

±1E+10

183

33/22

0xA1B0…0xA1B1/41393…41394

This is the capacity zero point for direct reading in sensor readout units.