beautypg.com

2 led functions, 4 basic rs232 operation, 1 hook-up – Bronkhorst LIQUI-FLOW mini (from 01-07-2013) User Manual

Page 14: Led functions, Basic rs232 operation, Hook-up, Section 3.4, Section 3.3.2

background image

LIQUI-FLOW™ mini

9.17.065

Bronkhorst High-Tech BV

14

LEDs

Time

Pushed

Indication

Green

Red

Off

Off

0 – 4 sec.

No action. Pressing a switch shortly by accident will not start any unwanted

reaction of the instrument.

Off

Normal

flash

0,2 sec on,

0,2 sec off

4 – 8 sec.

Restore factory settings

All parameter settings (except field bus settings) will be restored to situation of

final test at BHT production.

Normal

flash

0,2 sec on,

0,2 sec off

On

8 – 12 sec.

For FLOW-BUS only: install a free node-address on FLOW-BUS.

Normal

flash

0,2 sec on,

0,2 sec off

Normal

flash

0,2 sec on,

0,2 sec off

12 – 16 sec

Activate 'Configuration Mode'. The baud rate and bus type are set to 38k4 and

RS232 FLOW-BUS (Propar). The 'Configuration Mode' is recognized by the

green LED blinking 2 sec on, 0,1 sec off. The 'Configuration Mode' is

deactivated only after applying this micro-switch action again.

LED indications using micro-switch at power-up situation of an instrument

3.3.2

LED functions

The LEDs on top of the instrument can also be used for manual operation of some options. The green LED will indicate in what

mode the instrument is active. The red LED will indicate error/warning situations.

i

www

For details see “Manual interface: micro-switch and LEDs” in Operation Instructions Digital Instruments:

http://www.bronkhorst.com/files/downloads/

manuals_english/917023_operation_instructions_digital_instruments.pdf (document nr. 9.17.023, Chapter 11)

3.4

Basic RS232 operation

Digital operation adds a lot of extra features (compared to analog operation) to the instruments, such as:

Up to eight selectable fluids (if installed)
Direct reading at readout/control module or host computer
Testing and self diagnosis
Identification (serial number, model number, device type, user tag)
Adjustable minimal and maximal alarm limits
(Batch) counter

3.4.1

Hook-up

Connecting a LIQUI-FLOW

TM

mini instrument to a COM port of a

pc requires a special cable (7.03.426) which changes the

appropriate pin configuration. Optionally use an RS232 to USB2.0

converter (9.02.122) to connect to a USB port. Use the split cable

(Y-adapter 7.03.241) in combination with the Plug-in Power

Supply (7.03.424) for powering the instrument.

Instead of using a COM or USB port, it is also possible to connect

the RS232 pinning manually using the loose-end cable (7.03.419),

typically for connection to PLC or micro controller devices.

See section 2.7 for the Hook-up diagram.