Bronkhorst LIQUI-FLOW mini (from 01-07-2013) User Manual
Page 27

LIQUI-FLOW™ mini
© 2013 Bronkhorst High-Tech BV
Bronkhorst High-Tech BV
27
Fieldbus 2 address
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0...255
309
124/10
0xFC50/64593
Set the node address at any of the allowed values listed in the table in Section 3.5.3, e.g. '3'.
Fieldbus 2 baud rate
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0...1E+10
310
124/9
0xFC48...0xFC49/64585...64586
Set the baud rate at any of the allowed values listed in table in section Section 3.5.3, e.g. '19200'.
Fieldbus 2 medium
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0...1
311
124/11
0xFC58/64601
Select the communication medium. Parameter value 0 = RS232, 1 = RS485
Fieldbus 2 parity
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0...2
336
124/12
0xFC60/64609
Set the parity for Modbus communication. Parameter value 0 = None, 1 = Odd, 2 = Even.