beautypg.com

4 display filter, 5 alarm / status parameters, 6 counter parameters – Bronkhorst LIQUI-FLOW mini (from 01-07-2013) User Manual

Page 23: Display filter, Alarm / status parameters, Counter parameters

background image

LIQUI-FLOW™ mini

© 2013 Bronkhorst High-Tech BV

Bronkhorst High-Tech BV

23

4.1.4

Display filter

The output signal of a LIQUI-FLOW

TM

mini instrument (measured value) is filtered. The filter has dynamic behaviour: when a change

in sensor signal is detected, the measured value will be less filtered than when the sensor signal is constant and stable. There are

two filter constants: Static Display Factor and Dynamic Display Factor. These two factors can be transformed into time constants

using the following formula:

The measured value is filtered with a first order low pass filter with a filter time constant between the two ô values.

Dynamic Display Factor

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

RW

0…1.0

56

117/1

0xF508…0xF509/62729…62730

This value should not be changed.

Static Display Factor

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

RW

0…1.0

57

117/2

0xF510…0xF511/62737…62738

This value should not be changed.

CycleTime

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char

R

0…255

52

114/12

0x0E4C/3661

Note: The unit of parameter CycleTime is 10 ms. Example: value 0.2 means 2 ms

4.1.5

Alarm / Status parameters

i

www

See document nr. 9.17.023: Operational instructions for digital instruments, chapter 6.

http://www.bronkhorst.com/files/downloads/

manuals_english/917023_operation_instructions_digital_instruments.pdf

4.1.6

Counter parameters

i

www

See document nr. 9.17.023: Operational instructions for digital instruments, chapter 7.

http://www.bronkhorst.com/files/downloads/

manuals_english/917023_operation_instructions_digital_instruments.pdf