Page 9: Zeroing, Calibration, Supply pressure

background image

9.17.045

page 9

Warm-up time

Let the IQ

+

FLOW® warm-up for at least 30 minutes for best accuracy.

2.10 Zeroing

The zero point of each instrument is factory adjusted. If required the zero point can be re-adjusted over

RS232 or by means of using the micro-switch. Procedure for zeroing by-micro switch (not for pressure

meter/controller):

-

Warm-up, pressure up the system and fill the instrument according to the process conditions.

-

Make sure no flow is going through the instrument by closing valves near the instrument.

-

The setpoint must be zero.

-

Press micro-switch and hold it. After a short time the red LED will go ON and OFF, then the green

LED will go ON. At that moment release the micro-switch.

-

The zeroing procedure will start at that moment and the green LED will blink fast. The zeroing

procedure waits for a stable signal and saves the zero. If the signal is not stable, zeroing will take

long and the nearest point to zero is accepted. The procedure will take approximately 10 seconds.

-

When the indication is showing 0% signal and the green indication LED is burning continuously

again, the zeroing action was successful.

2.11 Calibration

Each IQ

+

FLOW® instrument is factory calibrated. Bronkhorst High-Tech B.V. certifies that all instruments

meet the rated accuracy. Calibration is performed using measurement standards traceable to the

standards of the Dutch Metrology Institute (VSL). Calibration certificates are included in the shipment.

When operated properly (clean gas, no pressure shocks, no vibrations, no thermal shocks, etc.), regular

maintenance is not required. However, periodical inspection, recalibration or verification of the

accuracy may be subject to individual requirements of the end-user.

2.12 Supply pressure

It is recommended to turn on power before applying pressure on the instrument and to switch off power

after removing pressure.

Turn on fluid supply gently. Avoid pressure shocks and bring the instrument gradually up to the level of the

actual operating conditions. Also switch off fluid supply gently.