4.2 Special instrument features

4.2.1 Auto zeroing

To start the auto zero-procedure by digital operation two parameters should be written:

Control Mode

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char

RW

0…255

12

1/4

0x0024/37

Calibration Mode

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char

RW

0…255

58

115/1

0x0E61/3682

Value

Mode

Instrument action

0

Idle

Idle

9

Auto zero

Auto-zeroing

255

Error

Idle

Auto-zero procedure:

Step 1: Set ‘Control Mode’ to ‘Calibration Mode’ (value 9)

Step 2: Set ‘Calibration Mode’ to ‘Auto zero’ (value 9)

Step 3: Check ‘Calibration Mode’,

Idle (value 0)

Auto-zeroing succeeded

Auto zero (value 9)

Auto-zeroing active

Error (value 255)

Auto-zeroing failed

4.2.2 Changing default control mode

Instruments are delivered with either analog or digital signal as default, depending on customer’s requirement. After

every (power-up) reset the instrument will return to its default control mode. The default control mode can be

changed with the following parameter:

IOStatus

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char

RW

0…255

86

114/11

0xF258…0xF259/62041…62042

Bit 6 [7…0] represents the former analog jumper.

1 = default control mode is analog

0 = default control mode is digital

Procedure for changing default digital operation to default analog operation:

Read ‘IOStatus’

Add 64 to the read value

])

40

0

[

(

x

OR

Write ‘IOStatus’

Procedure for changing default analog operation to default digital operation:

Read ‘IOStatus’

Subtract 64 from the read value

])

40

0

[

(

x

AND

Write ‘IOStatus’