5 leak check, 6 electrical connection, Leak check – Bronkhorst IQ+FLOW (till 01-07-2013) User Manual

Page 7: Electrical connection

background image

9.17.045

page 7

Compression type fittings

For leak tight installation of compression type fittings make sure that the tube is inserted up to the

shoulder in the fitting body and that no dirt or dust is present on tube, ferrules or fittings. Tighten the

nut finger tight; while holding the instrument, then tighten the nut one turn.

10-32 UNF fittings

Tighten the 10-32 UNF fittings according to the instructions

of the supplier of the fittings.

Only use 1/16” tubing with a straight and clean cut without

burrs to ensure leak tightness. Preferably deburr the tubing

prior to installation. A new ferrule connection must be made

for each new adapter to ensure leak-tightness and minimum

dead volume, due to variances in the adapter dimensions.

Mounting downported instruments

For downported instruments make sure that the seals are present at the bottom, that the surfaces are

undamaged and that they are dry and free from dirt or dust.

2.5 Leak check

Check the system for leaks before applying pressure. Especially if toxic, explosive or other dangerous fluids

are used!

2.6 Electrical connection

Electrical connections must be made with a standard cable or according to the IQ

+

FLOW® hook-up

diagram. Several hook-up examples and standard cables are found in chapter 3.

IQ

+

FLOW® instruments are powered with +15…+24 Vdc.

Electrical connections must be made with standard cables or according to the applicable hook-up

diagrams. These documents can be found at: http://www.bronkhorst.com/en/downloads

The instruments contain electronic components that are susceptible to damage by electrostatic discharge.

Proper handling procedures must be taken during installation, removing and connecting the electronics.

The instruments described in this manual carry the CE-mark and are complient with the EMC

requirements. However compliance with the EMC requirements is not possible without the

use of proper cables and connector/gland assemblies. Bronkhorst High-Tech B.V. recommends the use of

their standard cables. These cables have the right connectors and if loose ends are used, these will be

marked to prevent wrong connection. When using other cables, cable wire diameters should be sufficient

to carry the supply current and voltage losses must be kept as low as possible. When in doubt: contact

your distributor.

When connecting the system to other devices (e.g. to PLC), be sure that the integrity of the shielding is not

affected. Do not use unshielded wire terminals.