Bronkhorst IQ+FLOW (till 01-07-2013) User Manual

Page 12

background image

9.17.045

page 12

3.3 Manual interface: micro-switch, LEDs and rotary switches

In this section the manual instrument interfaces are described. In section 3.3.1 the micro-switch operation is discussed

(if present), section 3.3.2 specifies the LED functionality and in section 3.3.3 the rotary switch operation is described.

3.3.1 Micro-switch operation (single-channel versions only)

By means of manual operation of the micro push-button switch some important actions for the instrument can be

selected or started. These options are available in both analog and digital operation mode.

These functions are:

Reset alarm

Reset instrument (firmware program reset)

Auto-zeroing

Restore factory settings (in case of accidently changing of the settings)

Using digital RS232 or RS485 operation it is also possible to set:

Bus-address (only required for RS485)

Baudrate

The tables below describe the micro-switch functions that can be started in respectively normal operation mode and

during power-up:

LEDs

Time

Pushed

Indication

Green

Red

Off

Off

0 – 1 sec

No action. Pressing a switch shortly by accident will not

start any unwanted reaction of instrument.

Pressing the switch 3x briefly with intervals of max. 1 sec.

forces the instrument to indicate its bus-address/MAC-ID

and baud rate. Check chapter 3.5 for more details.

Off

Off

1 – 4 sec

In case of min/max alarm or counter batch reached:

Reset alarm (only if reset by keyboard has been enabled)

For FLOW-BUS only: if the node address is occupied, this

function will install a free node-address on FLOW-BUS.

Off

On

4 – 8 sec

Reset instrument

Instrument program will be restarted and all warning and

error messages will be cleared. During start-up the

instrument will perform a (new) self-test.

On

Off

8 – 12 sec

Auto-zero

Instrument will be re-adjusted for measurement of zero-

flow (not for pressure meter/controller). See section 2.10.

On

On

12 – 16 sec Prepare instrument for FLASH mode for firmware update.

Instrument shuts down and both LEDs turn off.

At next power-up instrument will be active again.

LED indications using micro-switch at normal operation mode of an instrument

LEDs

Time

Pushed

Indication

Green

Red

Off

Off

0 – 4 sec

No action. Pressing a switch shortly by accident will not

start any unwanted reaction of the instrument.

Off

Normal flash

0,2 sec on,

0,2 sec off

4 – 8 sec

Restore factory settings

All parameter settings (except field bus settings) will be

restored to situation of final test at BHT production

Normal

flash

0,2 sec on,

0,2 sec off

Off

8 – 12 sec

For FLOW-BUS only: install a free node-address on FLOW-

BUS.

Normal

flash

0,2 sec on,

0,2 sec off

Normal flash

0,2 sec on,

0,2 sec off

12 – 16 sec Activate “Configuration Mode”. The baud rate and bus

type are set to 38K4 and RS232 FLOW-BUS (Propar). The

“Configuration Mode” is deactivated only after applying this

micro-switch action again.

LED indications using micro-switch at power-up situation of an instrument

Multi-functional switch

Status LED

Error/Warning LED