Software

When using a pc to communicate with IQ

+

FLOW® instruments only the FLOW-BUS protocol is supported by Bronkhorst

software. When using Modbus operation, software from third parties, such as LabView, ModScan or a Modbus PLC must

be used to serve as Modbus master.

Note: an IQ

+

FLOW® instrument set for RS485 FLOW-BUS or Modbus communication will not react when

connecting to an RS232 configuration. When required press the micro-switch according to the procedure

in section 3.3.1 at start-up to activate the “Configuration Mode”. The baud rate and bus type are set to

38K4 and RS232 FLOW-BUS (Propar). For multi-channel instruments the communication type can be set

with the rotary switches.

Slave address, baud rate and parity setup

The IQ

+

FLOW® instrument is configured as specified on order. If there is a need of changing any of the specified

settings, see the tables below for the supported configurations.

Single-channel versions

Mode:

Analog

Digital

Interface:

-

RS232

RS485

Bus protocol:

-

Propar

FLOW-BUS

Modbus RTU

Modbus ASCII

Baud rate:

-

9600

19200

38400

57600

115200

187500

400000

9600

19200

38400

56000

57600

115200

9600

19200

38400

56000

57600

115200

Node address:

-

3…125

3…125

1…247

1…247

Parity:

-

None*

None*

None; Even; Odd

None; Even; Odd

Multi-channel versions (select with rotary switches)

Mode:

Digital

Medium:

RS232

RS485

Bus protocol:

Propar

Modbus RTU

Baud rate:

38400

115200

9600

19200

38400

Node address:

3…99

1…99

Parity:

None*

Even

* Not selectable