When using a pc to communicate with IQ
+
FLOW® instruments only the FLOW-BUS protocol is supported by Bronkhorst
software. When using Modbus operation, software from third parties, such as LabView, ModScan or a Modbus PLC must
be used to serve as Modbus master.
Note: an IQ
+
FLOW® instrument set for RS485 FLOW-BUS or Modbus communication will not react when
connecting to an RS232 configuration. When required press the micro-switch according to the procedure
in section 3.3.1 at start-up to activate the “Configuration Mode”. The baud rate and bus type are set to
38K4 and RS232 FLOW-BUS (Propar). For multi-channel instruments the communication type can be set
with the rotary switches.
Slave address, baud rate and parity setup
The IQ
+
FLOW® instrument is configured as specified on order. If there is a need of changing any of the specified
settings, see the tables below for the supported configurations.
Single-channel versions
Mode:
Analog
Digital
Interface:
-
RS232
RS485
Bus protocol:
-
Propar
FLOW-BUS
Modbus RTU
Modbus ASCII
Baud rate:
-
9600
19200
38400
57600
115200
187500
400000
9600
19200
38400
56000
57600
115200
9600
19200
38400
56000
57600
115200
Node address:
-
3…125
3…125
1…247
1…247
Parity:
-
None*
None*
None; Even; Odd
None; Even; Odd
Multi-channel versions (select with rotary switches)
Mode:
Digital
Medium:
RS232
RS485
Bus protocol:
Propar
Modbus RTU
Baud rate:
38400
115200
9600
19200
38400
Node address:
3…99
1…99
Parity:
None*
Even
* Not selectable