Operational instructions for digital multibus instruments

9.17.023

11.3.3 Change bus-address/MAC-ID and baudrate:

Pressing the switch 5x briefly with intervals of max. 1 second in normal running/operation mode. Within the time-out

period of 60 seconds it is possible to start changing the bus-address/MAC-ID of the instrument. For certain field bus

systems it is necessary to select the baud rate also. Other field bus systems only have one baud rate or the baud rate

setting will adapt to the setting of the master automatically. In these cases baud rate selection is not needed and will

be skipped.

Procedure for changing bus-address/MAC-ID and baud rate

step

action

Indication

time

handling

1

Start

Press the switch 5x briefly with intervals of max. 1

second in normal running/operation mode.

2

Set tens of bus-

address/MAC-ID

Green LED flashes

0.1 sec on

0.1 sec off

count-flashes

start when switch

is pressed:

0.5 sec on,

0.5 sec off

time-out:

60 sec

Press switch and count green flashes for tens of

bus-address/MAC-ID.

Release when wanted amount has been count.

Counts up to max. 12 and than starts at 0 again.

When counting fails, keep switch pressed and

restart counting for next attempt.

3

Set units of bus-

address/MAC-ID

red LED flashes

0.1 sec on,

0.1 sec off

count-flashes

start when switch

is pressed:

0.5 sec on,

0.5 sec off

time-out:

60 sec

Press switch and count red flashes for units of

bus-address/MAC-ID.

Release when wanted amount has been count.

Counts up to max. 9 and than starts at 0 again.

When counting failed, keep switch pressed and

restart counting for next attempt.

4

Set baud rate of field

bus communication.

Only for specific

types of field busses:

e.g. DeviceNet.

This part will be

skipped if no baud

rate needs to be

selected.

both red

and green

LED flashes

0.1 sec on,

0.1 sec off

count-flashes

start when switch

is pressed:

0.5 sec on,

0.5 sec off

time-out:

60 sec

Press switch and count red and green flashes for

baud rate setting of the specific field bus.

Release when wanted amount has been count.

Counts up to max. 10 and than starts at 0 again.

When counting failed, keep switch pressed and

restart counting for next attempt.

Note: selection of 0 means: No change

Instrument returns to normal running/operation mode. Changes are valid when they are made within the time-out

times.

Value zero will be indicated by a period of 1 sec. off (0.5 sec. off + 0.5 sec. off).

When value zero is wanted, press switch shortly and release it again within 1 sec.

Before each action of flash-counting, the LED’s to be used for counting will flash in a high frequency.

(Pattern: 0.1 sec on, 0.1 sec off). As soon as the switch is pressed-down, this LED (or both LED’s) will be

off and the counting sequence will start.