Operational instructions for digital multibus instruments

9.17.023

11.2 LED’

S INDICATIONS

11.2.1 LED indications mode (no switch used)

Led

Time

Indication

Green

off

Continuous Power-off or program not running

on

Continuous Normal running/operation mode

Short

flash

0.1 sec on

2.0 sec off

Initialization mode (Init reset = 73)

For MBC3 type: no bus communication, safe state active.

normal

flash

0.2 sec on

0.2 sec off

Special function mode

Instrument is busy performing any special function. E.g. auto-zero or self-test

long flash 2.0 sec on

0.1 sec off

Special mode, see specific field bus for more details
FLOW-BUS

Not used

PROFIBUS-DP /Modbus Not used
DeviceNet (MBC-II)

Idle state

DeviceNet (MBC3)

See special table below

EtherCAT

Not used

Red

off

Continuous No error

Short

flash

0.1 sec on

2.0 sec off

Special mode, see specific field bus for more details
FLOW-BUS

Node occupied: Re-install instrument

PROFIBUS-DP

No data-exchange between master and slave Automatic recovery

Modbus

Data is received or transmitted

DeviceNet (MBC-II)

Minor communication error

DeviceNet (MBC3)

See special table below

EtherCAT

Instrument is not in OP mode (see EtherCAT manual for details)

normal

flash

0.2 sec on

0.2 sec off

Warning message.

An error occurred of minor importance.

It would be wise to investigate the cause of this.

You are still able to work with your instrument.

See specific field bus for more details
FLOW-BUS

Waiting for communication

PROFIBUS-DP / Modbus No details
DeviceNet (MBC-II)

No bus power

DeviceNet (MBC3)

See special table below

EtherCAT

Not used

long flash 2.0 sec on

0.1 sec off

See specific field bus for more details
FLOW-BUS

Not used

PROFIBUS-DP

A requested parameter is not available.

See troubleshoot in Profibus manual.

Modbus

For special service purpose only

DeviceNet (MBC-II)

Serious communication error; manual intervention needed

DeviceNet (MBC3)

See special table below

EtherCAT

Error detected in EtherCAT configuration (see EtherCAT manual

for details)

on

Continuous Critical error message. A serious error occurred in the instrument.

Instrument needs service before further using.

Wink Mode Green Red Green Red turn by turn
slow

wink

0.2 sec on

0.2 sec off

Wink mode

By a command send via FLOW-BUS the instrument can “wink” with Led’s to indicate its

position in a (large) system

normal

wink

1.0 sec on

1.0 sec off

Alarm indication: minimum alarm, limit/maximum alarm; power-up alarm or limit exceeded

or batch reached.

fast

wink

0.1 sec on

0.1 sec off

Switch-released, selected action started