Operational instructions for digital multibus instruments
9.17.023
7.7 C
OUNTER UNIT
unsigned char[4]
string
RW
N
128
104/7
This parameter gives access to the extended counter unit table which is available for MBC3 type of
instruments only.
This parameter can only be read for MBC-II type of instruments.
For MBC3 type of instruments this parameter can be read and written.
The easiest way to change a unit in the MBC3 type of instrument is to fill in the unit needed from the
table below.
The “Counter unit” displays the unit name set by “Counter unit index”. A valid “Counter unit”(for example ln) can also be
entered here which changes the “Counter unit index”.
In MBC3 type of instruments the parameter is not secured.
Extended counter unit table
Mass
ug
mg
g
kg
Custom volume
ul
ml
l
mm3 cm3 dm3
m3
Normal volume
uln
mln
ln
mm3n cm3n dm3n m3n
Standard volume
uls
mls
ls
mm3s cm3s dm3s m3s
7.8 R
ESET COUNTER ENABLE
Unsigned char
0…15
RW
N
157
104/9
Available counter reset options:
Automatic
Reset
par 114
External*
Keyboard/
micro-switch
Value
bit[3]
bit[2]
bit[1]
bit[0]
0
0
0
0
0
1
0
0
0
1
2
0
0
1
0
3
0
0
1
1
4
0
1
0
0
5
0
1
0
1
6
0
1
1
0
7
0
1
1
1
8
1
0
0
0
9
1
0
0
1
10
1
0
1
0
11
1
0
1
1
12
1
1
0
0
13
1
1
0
1
14
1
1
1
0
15
1
1
1
1
*External is not used in MBC-II and MBC3 type instruments.