Operational instructions for digital multibus instruments
9.17.023
11.3.2 Readout bus-address/MAC-ID and baudrate:
Pressing the switch 3x briefly with intervals of max. 1 second in normal running/operation mode will trigger the
instrument to “show” its bus address/MAC-ID and baud rate. For indication the bus-address/MAC-ID the green LED
will flash the amount of tens and the red LED the amount of units in the number. For indication of baud rate setting,
both LED’s will flash. The flashes are called “count-flashes” and have a pattern of 0.5 sec. on, 0.5 sec. off.
LED indications for bus-address/MAC-ID and baud rate (press switch 3x briefly)
LED
LED
Time
Indication
Green
Red
amount of count flashes
(0...12)
Off
0 ... 12 sec.
Maximum
tens in bus-address/MAC-ID for instrument
off
Amount of count flashes
(0...9)
0 ... 9 sec.
Maximum
units in bus-address/MAC-ID for instrument
amount of count flashes
(0...10)
amount of count flashes
(0...10)
0 ... 10 sec.
Maximum
baud rate setting for instrument
Value zero will be indicated by a period of 1 sec. off (0.5 sec. off + 0.5 sec. off).
Examples:
• For bus address/MAC-ID 35 the green LED will flash 3 times and the red LED will flash 5 times.
• For bus address/MAC-ID 20 the green LED will flash 2 times and the red LED will flash 0 times.
• For bus address/MAC-ID 3 the green LED will flash 0 times and the red LED will flash 3 times.
• For bus address 126 the green LED will flash 12 times and the red LED will flash 6 times.
Baud rate index table for indication on the LED’s (in baud)
FLOW-BUS
PROFIBUS-DP
DeviceNet
Modbus
EtherCAT
1
2
187500
400000*
0
1
2
3
4
5
6
7
8
9
10
not detected
9600
19200
45450
93750
187500
500000
1500000
3000000
6000000
12000000
1
2
3
125000
250000
500000
1
2
3
4
5
9600
19200
38400
57600*
115200*
1
100000000
*MBC3 type instruments have additional baud rates available for the several field busses.
EtherCAT bus address is always ‘0’.
Examples:
• For PROFIBUS-DP baud rate readout of 12000000 Baud, both LED’s will flash 10 times.
• For DeviceNet baud rate readout of 250000 Baud, both LED’s will flash 2 times.