Operation
Sorensen SLH-Series DC High Power Electronic Load
Shorts Mode
1. Shorts mode puts the load in minimum resistance/maximum current state.
2. To turn on shorts mode, press the SHORT key (11) until the LED turns on. This
overrides any selection using the MODE key. When the LOAD key is toggled to ON,
the load simulates a short by sinking maximum rated current at minimum resistance.
3. To exit shorts mode, press the SHORT key until the LED turns off.
Normal Operating Mode Flow Chart:
MODE
Press MODE
Button until
desired mode is
illuminated CC,
CR, CV, CP
PRES
LOW
HIGH
DYN/STA
LEVEL
SHORT
LOAD
T
HIGH
T
LOW
FALL
RISE
Dynamic
Shorts MODE?
YES
NO
If only one value is
desired, set HIGH
to maximum rated
value. Use LOW
to set desire value.