beautypg.com

Current -2, Measurement -2, 2 current – AMETEK CW-M User Manual

Page 12: 3 measurement

background image

Elgar CW-M Series

2-2

M161570-01

Load Regulation: ±0.1% of full scale voltage for a full resistive load to no load

Line Regulation: ±0.1% of full scale for a ±10% line change from nominal line

voltage

Remote Voltage Sense: 5 VRMS total lead voltage drop

2.2 CURRENT

CW 801M: 6.00 ARMS in 135 VAC range or 3.00 ARMS in 270 VAC range

CW 1251M: 9.40 ARMS in 135 VAC range or 4.70 ARMS in 270 VAC range

CW 2501M: 18.60 ARMS in 135 VAC range or 9.30 ARMS in 270 VAC range

Frequency

Range: 45 to 500 Hz

Accuracy: ±0.5% typical

Resolution: 0.1 Hz

2.3 MEASUREMENT

Voltage

Range: 0 to 135/270 VRMS

Accuracy: ±1% of range for voltages >5V, measured at point of sense

Resolution of Display: 0.1V

Current

Range: 0 to 6.0 ARMS (CW 801M)

0 to 9.4 ARMS (CW 1251M)
0 to 18.6 ARMS (CW 2501M)

Accuracy: ±2% of range for linear loads with current >0.2A for CW 801M and

CW 1251M, >0.4A for CW 2501M

Resolution of Display: 0.1A

Frequency

Range: 45 to 500 Hz

Accuracy: ±0.5% of reading

Resolution of display: 0.1 Hz