Dataman 48UXP User Manual
Page 68
Device inserted backwards or damaged already!
The Dataman-48UXP detected that the device has been inserted the
wrong way round, or that the device has been damaged.
Verify that the device has been inserted correctly. If not, try another
device to verify that you are using the correct algorithm etc. to
program the device.
Device inserted upwards %d slot(s)!
The Dataman-48UXP has detected that the device has been inserted
incorrectly. Remove the device and insert it correctly.
Device insertion error or damaged already!
The Dataman-48UXP detected a device in socket, but several pins
do not respond as expected (poor contact). This indicates that a
device has been inserted incorrectly or that the device has been
damaged. Verify that the device has been inserted correctly. If not,
try another device to verify that you are using the correct algorithm
etc. to program the device.
File write error!
The Dataman-48UXP detected an error when writing a file to disk.
Check that there is enough space on the disk to hold the file. Also
check that the disk is not write-protected.
This might happen on a network if you are a user that does not have
rights to the directory you want to save the file to. Use another
directory or disk.
Function code has not been initialized yet!
This message indicates that the data file is corrupt. Suggest
re-install the Dataman-48UXP software and remove all the
Dataman-48UXP old file in your disk before installation.