Ntellisys, Daptive, Ressure – Nor-Cal IQ+ Adaptive Pressure Controller User Manual
Page 17: Ontrollers, 1 - how to configure iq+ for your system, Devicenet commands, Continued), System configuration commands
1
Call toll free
800-84-4166
or 530-842-4457
•
FAX 530-842-9130
I
ntellIsys
A
dAptIve
p
ressure
C
ontrollers
IQ+ OP-LIT 9/08
6.1 - How to Configure IQ+ for your System
(continued)
FIGURE 6.4 – SPEED TOO HIGH
FIGURE 6.5 – SPEED TOO LOW
System Configuration Commands
DeviceNet Commands
Class ID 33
hex
Instance ID 1 (Process Control)
attrIBute Id hex
acceSS code
name
Format
value
64
OE (get) 10 (set)
Volume
INT
1-100, default = 0
65
OE (get) 10 (set)
Delay
INT
0-10, default = 0
66
OE (get) 10 (set)
Speet
INT
1-100, default = 100
TAbLE 6.4 – DEvICENET COmmANDS
TAbLE 6.5 – RS-232 SERIAL COmmANDS
SerIal command
deScrIPtIon
reSPonSe
SVxxx
Set Volume where xxx is a number from 1 to 100
PID VOLUME: xxx
SDxxx
Set Delay where xxx is a number from 1 to 10
PID DELAY: xxx
SSxxx
Set Speed where xxx is a number from 1 to 100
PID SPEED: xxx
TAbLE 6.6 – RS-232 SERIAL REQUESTS AND RESPONSES
SerIal requeSt
deScrIPtIon
reSPonSe
RV
Requests Volume value
PIC VOLUME: xxx
RD
Requests Delay value
PID DELAY: xxx
RS
Requests Speed value
PID SPEED: xxx
RPI
Requests values of Volume, Delay and Speed
SPEED: xxx
VOLUME: xxx
DELAY: xxx
A
Record the current Volume, Delay, and Speed parameters.
b
For serial interface set the serial port speed 100ms or less. For Devicenet
set to 50ms. Or less.
C
Start pressure control at a process critical set point, or a set point halfway
between the system maximum and minimum process pressure set points.
With the system GUI or a PC that can graph data observe the pressure and
valve movement.
D
If the pressure parameters need adjusting stop pressure control and Open
the valve. Enter new value for pressure parameters.
Parameters are adjusted in this order: volume
Delay
Speed (not normally adjusted)
E
If the pressure response is satisfactory, input a set point at the system
minimum pressure or another process critical set point. Make adjustments
as necessary. If a change was made re-check pressure set point in step C.
F
If the pressure response is satisfactory, input a set point at system maxi-
mum pressure. Adjust if necessary. If a change was made re-check set
points in steps C and D.
G
Record new parameters for Volume, Delay, and Speed.
Setting Volume, Delay and Speed
Setting Volume, Delay and Speed (Example)
The system has a 1 Torr pressure gauge. The pressure control range is
20mTorr - 600 mTorr. The process critical range is 120mTorr.
A
Record values of Volume, Delay, and Speed.
b
Start pressure control with a set point 120mTorr. Adjust the parameters if
necessary.
C
Start pressure control with a set point of 20mTorr. Adjust parameters if
necessary. If any changes are made then re-check the 120mTorr set point.
D
Start pressure control at 600mTorr.. Adjust the parameters if neces-
sary. If any changes are made then re-check the 120mTorr and 20mTorr set
points.
E
Record new values of Volume, Delay, and Speed.