Laerdal SimMan Essential User Manual
Page 53

53
SETUP FOR SIMULATION
Cleaning the genitalia pad
1 Use a syringe and flush the genitalia with approximately 200ml of water through the genitalia opening.
2 Press the bulb on the back side of the genitalia pad to expel the last of the water.
3 Let all the parts air dry, or wipe them with a cloth to remove excess water.
4 Ensure that the urine tank is empty of water before storing the manikin.
Cleaning FOR StORage OR eXtended peRiOdS OF nOn-USe
1 Fill the urine system with 60% Isopropanol or 70% Ethanol
2 Insert a catheter, and drain the system
3 Remove catheter
4 Connect a syringe with 100ml of air, and fill the urine system with air
5 Insert a catheter and drain the urine system
6 Remove catheter
7 Repeat steps from 4. until the 60% Isopropanol or 70% Ethanol has been drained out and only air
exits the catheter.
ReplaCe URine tank
2 Open the stomach as shown in section
Opening the Torso – Manikin Setup.
3 Disconnect the tube labeled ‘P’ remove the urine tank
from the pelvis.
4 Put the new urine tank back into the manikin and
connect the corresponding connectors.
1 Remove the genitalia pad clear of the pelvis to reduce
the risk of leakage inside the manikin. Disconnect
the exposed tube connector labelled ‘G’.