CONFINED SPACE MODE
CONFINED SPACE MODE
This mode is used for confined space pre-entry testing
and for personal monitoring in areas such as basements
etc.
5.1 CSM RANGES
Confined Space Mode (CSM) will have the following
ranges available:
0 100% LEL (EEE Over-range)
0 25% O
2
(if fitted)
0 2000 ppm CO (if fitted)
0 100 ppm H
2
S (if fitted)
5.2 CSM FEATURES
Confined Space mode will have the following features
available:
Audible / Visual Alarms
Min / Max Display
Backlight
Flashlight
Pump ON continuously - will not stop if flow fault
is detected - flow fault alarm will be activated.
Datalogging automatic every 60 seconds