USER HANDBOOK
4) Inspect the filter for contamination or damage.
If either is evident, replace the filter.
5) Fit a new Hydrophobic Probe Filter (Part No.
67259), if required.
Note:
Installation is a reversal of the removal
procedure and therefore relevant Figs.
should be referred to.
6) Position hydrophobic probe filter so that slots in
filter locate corectly over lugs in probe tube then
push filter until correctly seated. Care must be
taken not to damage the figaro sensor.
7) Hold probe tube then attach probe filter bulb to the
flexible probe assembly, in a clockwise direction,
and tighten to secure.
Note:
Care must be taken not to overtighten the
probe filter bulb.
8) Before attaching the probe connector to the probe
filter bulb, check the probe washer for signs of
damage. If damaged, it must be replaced with a
new probe washer (Part No.12379).
9) Hold probe filter bulb then attach probe connector
to the probe filter bulb, in a clockwise direction,
and tighten to secure.
Note:
Care must be taken not to overtighten the
probe connector.