PRINTING/LOGGING WITH HI933300

To print the measured values
shown on display, press the
PRINT key once.

This function can be activated in normal op-
eration mode as well as during logging and
scanning data on display (see below).

When in measurement mode, the printout pro-
vides the following information:

a - Running log number

b - Date (DD-MM-YY)

c - Time (HH-MM)

d - Conductivity value in µS/cm or mS/cm

e - Temperature in degrees Centigrade.

LOGGING MODE

This function is suggested when remote mea-
surements have to be taken automatically
without the necessity of an operator and for a
long period of time. In this mode data will be
stored directly into memory.

Set the appropriate logging interval (see Op-
erational Guide section on page 8).

Press the ALT and the LOG keys simulta-
neously to enter the logging mode. The log
number and page number will appear for a
few seconds on the display to indicate the
correct operational mode. The printer will print
a complete set of data and the "LOG" sym

PRINT

2 9 - 0 3 - 9 6

- - 0 1 - -

* 1 8 . 0 1

1 4 5 5 µ S

2 5 . 9 ° C

a

b

c

d

e