PROCEDURE FOR OFFSET CALIBRATION
(at 0.0 µS/cm)

To perform the Offset

calibration, dry the con-
ductivity probe and leave
it in air.

Press the ALT and the CAL keys simulta-

neously. The primary LCD display will blink
"

0.0

µS" for about 10 seconds.

When "

0.0

µS" stops blinking, the cali-

bration can be confirmed. Press the CFM
key to confirm the offset value.

If the input reading is out of the offset

range, an error message will show on the
secondary display "

E0

". Press the CAL

key to re-calibrate the value.

If everything is satisfactory

the LCD will blink
"

84.0

µS", expecting the

first calibration solution.

Press the CAL key to exit

the calibration mode now.