When the buffer value stops blinking, indi-

cating that the probe has stabilized, press
the CFM key to confirm the value.

Press the CAL key to exit the calibration

mode now.

P R O C E D U R E F O R A C O M P L E T E
CALIBRATION WITH 5 BUFFERS

HI933300 and HI933301 have 5 memorized
calibration values:

0.0 µS/cm (probe in air)

84.0 µS/cm (cal. solution HI 7033/HI8033)
1413 µS/cm (cal. solution HI 7031/HI8031)
12.88 mS/cm (cal. solution HI 7030/HI8030)
80.0 mS/cm (cal. solution HI 7034/HI8034)

These meters can be calibrated at all the
above five points as described below.

To perform the Offset cali-

bration, dry the conductiv-
ity probe and leave it in
air.

Press the ALT and the CAL keys simulta-

neously. The primary LCD display will blink
"

0.0

µS" for about 10 seconds.

mS

CFM

CAL

CAL

ALT

µ

S