29

If everything is satisfactory

the LCD will blink "

50

"

expecting the second
buffer (50°C/122°F).

Note: if a single point cali-

bration is required,
press the CAL key
to leave the calibra-
tion mode now.

Place the probe into the

second container with hot
water (50°C/122°F).

When "

50

" stops blinking, press the CFM

key again to confirm the second calibra-
tion point.

If the value is out of the slope range, an

error message will show on the secondary
display "

E2

". Press the CAL key to re-

calibrate the value.

If everything is satisfactory the tempera-

ture calibration is complete and the meter
will return to the normal operational mode.

CAL

CFM

CAL

°C

50 °C

(122 °F)