into the PVC sleeve with the holes towards

the top (the end nearest to the cable) of
the probe.

Place the probe into the

container with ice and wa-
ter (0.0°C/32°F).

Press the CAL key. The secondary (lower)

display will blink "

0

" for about 30 sec-

onds.

When "

0

" stops blinking, the calibration

can be confirmed. Press the CFM key to
confirm the first value.

If the value is out of the offset range, an

error message will show on the secondary
display "

E1

". Press the CAL key to re-

calibrate the value.

CAL

CFM

CAL

0 °C

(32 °F)

°C