27

TEMPERATURE CALIBRATION

(for Technical Personnel only)

INITIAL PREPARATION

Prepare a container of ice and water and
another container with hot water (at a tem-
perature of at least 50°C/122°F). Place insu-
lation material around the containers to mini-
mize temperature changes.

Use a ChecktempC or a calibrated thermom-
eter with a resolution of 0.1° as reference
thermometer.

Note: After performing the following calibra-

tion, conductivity must be recalibrated.

PROCEDURE

Press the RANGE key to select the tem-

perature measurement mode (just tem-
perature measurement on the primary dis-
play).

Ensure that the probe is connected to the

meter securely by aligning the pins with
the socket, pushing the plug in and tight-
ening the threaded ring. Insert the probe

0 °C

(32 °F)

°C

50 °C

(122 °F)

°C

RANGE

°C