TAG INSTALLATION
The tag is housed in a rugged metal that can withstand harsh environments. However, it is better
to protect the tag from direct rain.
Place the tag near a sampling point. Fix it securely with the provided screws, in such a way that
the metallic iButton
®
is easily accessible for reading the tag.
The number of tags that can be installed is practically unlimited. Additional tags can be ordered
(
HI 920005 - five tag holders with tags).
LAMP REPLACEMENT
The instrument tungsten lamp has a life longer than 100,000 measurements. In case of lamp
failure, the defective lamp can be easily replaced. When the lamp is broken, the instrument displays
“
no L” error message.
To replace the lamp follow the next steps:
• Remove the battery lid.
• Unscrew the lamp connection using a
screwdriver.
• Unlock the lamp and extract it by pulling it out
from the lamp holder handler.
• Place the new lamp in the right position and
push it until is securely locked.
• Insert the lamp leads into the connector and
tight them using a screwdriver.
Warning: After lamp replacement the meter has
to be recalibrated.