• Press and hold

CAL/CHECK for 3 seconds to

enter calibration. The LCD will show “

CAL”

tag and the parameter for which the calibration
is performed.

• Place the CAL CHECK™ Standard Cuvette A into

the holder and ensure that the mark on the
cuvette is aligned with the mark on the
instrument top.

• Close the lid and press

ZERO/AVG

. The

LCD will display blinking dashes and the
measuring icon during zero measurement. At
the end of the zero measurement, “

-0.0-” is

displayed. The “

READ” tag will blink.

• Remove the cuvette.

• Place the CAL CHECK™ Standard Cuvette B into

the holder. Make sure that the mark on the
cuvette is aligned with the mark on the
instrument top.

• Close the lid and press

READ/TIMER . The

instrument will show blinking dashes and the
measuring icon during measurement.

• At the end, the value of the CAL CHECK™

standard value (1.00 mg/L) is displayed for
one second and then “

Stor” to confirm that the

new calibration data has been accepted.

• The meter automatically enters in measurement

mode.