SINGLE SAMPLE READ
• Remove the cuvette from the instrument.
• Remove the cap.
• Add the content of one packet of the specific
test reagent, for:
Free Chlorine
Total Chlorine
1 packet of or 1 packet of
HI 93701-0
HI 93711-0
• Replace the cap and shake gently for 20 seconds
(or 2minutes in case of seawater analysis).
• Replace the cuvette into the holder and ensure
that the mark on the glass is matched with the
mark on the instrument top. Close the lid.
• Hold
READ/TIMER for 3 seconds. The
display will show the hourglass blinking and
the countdown prior to measurement.
Alternatively wait for:
Free Chlorine
Total Chlorine
1 minute or 2 minutes and 30 seconds
and then just press
READ/TIMER .
• In both cases blinking dashes will appear during
measurement.
• The instrument directly displays the concentration
in mg/L of free or total chlorine.
Note: If the value is over range, the maximum
value (5.00 mg/L) will blink.
or
- pH 213 (17 pages)
- pH 211R (13 pages)
- pH 21 (9 pages)
- pH 209V (9 pages)
- pH 209R (13 pages)
- HI 99181 (2 pages)
- HI 99171 (2 pages)
- HI 99163N (2 pages)
- HI 99161N (2 pages)
- HI 99141 (2 pages)
- HI 99131 (2 pages)
- HI 99121 (9 pages)
- HI 99104 (2 pages)
- HI 991003 (2 pages)
- HI 991002 (2 pages)
- HI 991001 (2 pages)
- HI 98402 (9 pages)
- HI 98401 (8 pages)
- HI 98308 (2 pages)
- HI 98240 (21 pages)
- HI 98203 (2 pages)
- HI 98202 (2 pages)
- HI 98201 (2 pages)
- HI 98185 (72 pages)
- HI 98172 (52 pages)
- HI 98160N (48 pages)
- HI 98160 (48 pages)
- HI 98150 (17 pages)
- HI 9815 (2 pages)
- HI 981410 (2 pages)
- HI 981408 (2 pages)
- HI 98121 (2 pages)
- HI 98120 (2 pages)
- HI 98113 (2 pages)
- HI 98110 (2 pages)
- HI 98108 (2 pages)
- HI 98107 (2 pages)
- HI 98106 (2 pages)
- HI 98103 (2 pages)
- HI 96108 (2 pages)
- HI 96107 (2 pages)
- HI 96106 (2 pages)
- HI 96104 (28 pages)
- HI 96101 (36 pages)
- HI 9321 (36 pages)