• Close the lid and press

READ/TIMER . The

display will show blinking dashes and the icons
for cuvette, detectors and lamp will appear
during measurement.

• At the end of the measurement, the third

calibration point (100 NTU) is displayed on
the primary LCD and “

CAL P.3” on the

secondary LCD and “

READ” tag will blink.

• At this moment it is possible to exit calibration

by pressing

CAL/CHECK. The instrument will

memorize the two-point calibration data and
will return to measurement mode.

THREE-POINT CALIBRATION

To perform a three-point calibration, continue the
procedure with the following steps:

• Remove the second standard cuvette.

• Place the 100 NTU standard cuvette (or the

third prepared formazin standard) into the
holder, with the cuvette mark aligned with the
sign on the instrument top.

• Close the lid and press

READ/TIMER . The

display will show blinking dashes and the icons
for cuvette, detectors and lamp will appear
during measurement.

• At the end of the measurement, the fourth

calibration point (750 NTU) is displayed on
the primary LCD and “

CAL P.4” on the

secondary LCD and “

READ” tag will blink.

• At this moment it is possible to exit calibration

by pressing

CAL/CHECK. The instrument will

memorize the three-point calibration data and
will return to measurement mode.