• Turn the instrument on by pressing

ON/OFF.

Make sure that the instrument is in the free or
total chlorine range (the desired one).

• Press

RANGE

to select the desired range

(free or total chlorine). When dashes appear
on the LCD, the instrument is ready. The

ZERO” tag will blink on the LCD.

• Place the zero cuvette (A) in the instrument

with the mark aligned with the mark on the
instrument top.

• Close the lid and press

ZERO/AVG

. The

LCD will display blinking dashes and the
measuring icon during zero measurement. At
the end of zero measurement the “

-0.0-” is

displayed. The meter is now ready for
validation.

COLORIMETER CALIBRATION

The

HI 93414 free and total chlorine colorimeter has a powerful CAL CHECK function that allows the

user to check the instrument calibration against a NIST traceable standard before making a set of
measurements. With the same standard, the instrument could be re-calibrated, if necessary.
Note: Free and total chlorine must be calibrated separately. Calibration of one range will not

calibrate the other range.

VALIDATION PROCEDURE
Warning: Do not validate or calibrate the instrument with standard solutions other than Hanna CAL

CHECK™ Standards, otherwise erroneous results will be obtained. For accurate validation
and calibration please perform test at room temperature, 18 to 25 ºC (64.5 to 77.0 ºF).