Hanna Instruments HI 93414 User Manual

Page 12

background image

12

SPECIFICATIONS

Turbidity

Range

0.00 to 9.99; 10.0 to 99.9 and 100 to 1000 NTU

Range selection

Automatically

Resolution

0.01 NTU from 0.00 to 9.99 NTU;
0.1 NTU from 10.0 to 99.9 NTU;
1 NTU from 100 to 1000 NTU

Accuracy

±2% of reading plus 0.02 NTU

Repeatibility

±1% of reading or 0.02 NTU, whichever is greater

Stray Light

< 0.02 NTU

Light Detector

Silicon Photocell

Method

Ratio Nephelometric Method (90°), ratio of scatter and
transmitted light; Adaptation of the

USEPA Method 180.1

and

Standard Method 2130 B.

Measuring mode

Normal, Average, Continuous.

Turbidity Standards

<0.1, 15, 100 and 750 NTU

Calibration

Two, three or four-point calibration

Free and total Chlorine

Range Free Cl

2

0.00 to 5.00 mg/L

Total Cl

2

0.00 to 5.00 mg/L

Resolution

0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 above 3.50 mg/L

Accuracy

±0.02 mg/L @ 1.00 mg/L

Detector

Silicon photocell with 525 nm narrow band interference filters

Method

Adaptation of the

USEPA Method 330.5 and Standard Method

4500-Cl G. The reaction between chlorine and DPD reagent
causes a pink tint in the sample.

Standards

1 mg/L free chlorine, 1 mg/L total chlorine

Calibration

One-point calibration