TWO-POINT CALIBRATION

• Turn the instrument ON by pressing

ON/OFF.

If you are not in turbidity range, first select the
range. If you are in turbidity range, when dashes
are displayed on the LCD, the instrument is
ready. The current time appears on the
secondary LCD, if selected in SETUP menu or

turb”, if the time is not displayed.

• Enter calibration mode by pressing

CAL/

CHECK for 3 seconds. The display shows “CAL
P.1”on the secondary LCD and no suggested
value and “

READ” tag will blink.

• If the prepared formazin is used, edit the

displayed value by pressing

or

keys

until the display shows the correct value.

• Place the <0.1 NTU standard cuvette (or the

prepared one) into the holder and ensure that
the cuvette mark is aligned with the sign on
the instrument top.
Note: Alternatively, press CFM to skip the

first calibration point.

• Close the lid and press

READ/TIMER . The

display will show blinking dashes and the icons
for cuvette, detectors and lamp will appear
during measurement.

• At the end of the measurement, the second

calibration point (15 NTU) is displayed on the
primary LCD and “

CAL P.2” on the secondary

LCD, and “

READ” tag will blink.

• Remove the first standard cuvette.

• Place the 15 NTU standard cuvette (or the

second prepared standard) into the holder, with
the cuvette mark aligned with the sign on the
instrument top.